Instrument Database

Other

HiPE-LAB

MAPEX Instrument Database Logo

General information

  • Investigation area
  • Techniques
    Power electronics testing
  • Manufacturer
    IALB
  • Fabrication year
    2021
  • Measured quantity
    Temperature, relative humidity, airflow, current, voltage, power
  • Main application
    Multimodal lifetime tests (under climate and electrical conditions, destructive and non destructive) on power electronics.
  • In-situ, real-time compatible
    Yes

Instrument specification

  • Technical aspects

    Electrical load system: Power range up to 10 MVA with a voltage range from 0 to 1,000 V AC, currents up to 9,000 A AC and a frequency range from 0 to 70 Hz.

    Climate room: Temperature range from -40 °C  to +120 °C, 10 % to 95 % relative humidity at +10 °C to +95 °C, Test room (l x w x h) 7.5m x 5.3 m x 4.3 m, Cooling water supply temperature for the DUT from +16 °C to +60 °C.

    Measurement system: Up to 72 channels with  2 MSamples/s for voltages and currents and up to 64 channels with 20 kSamples/s temperature, relative humidty and airflow. Fully automated tests due to an flexible and extendable controll system.

  • In-situ capabilities
    Temperature range from -40 °C to +120 °C; relative humidity 10 % to 95 % at +10 °C to +95 °C; Cooling water supply temperature for the DUT from +16 °C to +60 °C.

Contact

Updated by: MAPEX